HCEM HC280A twin-head oil-free piston vacuum pump with finned cylinder heads

How to Measure Vacuum Pressure: Gauge Types, Accuracy and Where to Fit the Gauge

HCEM Engineering Team

To measure vacuum, connect a vacuum gauge where the pressure matters and read it on the right reference. In rough vacuum (from atmosphere down to about 1 mbar), where oil-free piston pumps like ours work, use a gauge that measures the force of the gas on a surface: a Bourdon dial for quick checks against the atmosphere, or a diaphragm gauge (capsule, piezo-resistive or capacitance) for accurate or absolute readings. These read the same whatever the gas is. Below about 10 mbar, thermal-conductivity (Pirani) gauges take over, and below about 0.001 mbar ionization gauges; both read differently for different gases. Before you trust any number, check three things: is it gauge or absolute, how accurate is the gauge at that end of its scale, and is the gauge fitted where the pressure really is.

This guide is for engineers who build vacuum into equipment, and for anyone who has to put a number on a vacuum: to accept a machine, to compare pumps, or to find out why a process misbehaves. It covers what you are actually measuring, the gauge types and their ranges, how accurate a dial really is near full vacuum (with a worked example), where to fit the gauge, and what to write down. Unit conversions are covered in our guide to gauge vs absolute pressure, and the step-by-step pump check in how to test a vacuum pump; this article is about the measurement itself.

Step 1: decide what you are measuring, gauge or absolute

A vacuum can be stated from either end of the scale. A gauge (relative) reading is the difference between the atmosphere and the vacuum, so it is negative or written as "vacuum": −90 kPa, 26.6 inHg. An absolute reading counts up from zero pressure: 113 mbar abs, 85 Torr. The two describe the same state only on a given day at a given place, because the atmosphere itself moves.

Which one you need depends on what the vacuum does in your application:

  • Gauge, when the job depends on the difference to atmosphere. A suction cup, a vacuum clamp or a vacuum lifter is pushed by the outside air, so its force follows the gauge vacuum. Validyne (read 2026-11-21) recommends measuring vacuum with a differential transducer that has one port open to atmosphere, because it then shows the vacuum correctly even when the local atmospheric pressure changes with the weather. That is exactly the quantity a cup or clamp depends on.
  • Absolute, when the process depends on the pressure itself. Processes such as degassing and drying, and pump-down calculations, depend on the absolute pressure, and so does any comparison of two pumps measured on different days or at different places. Fluid Power Journal (read 2026-11-21) recommends an absolute unit such as Torr or mbar(a) wherever a known vacuum level is needed, and calls inches of mercury a guide for applications such as vacuum cups.

How much does the atmosphere move? Fluid Power Journal quotes a sea-level high of about 15.6 psi, a low of about 12.5 psi inside a hurricane, and an average of about 12.1 psi in Denver. The US National Weather Service (read 2026-11-21) gives a rule of thumb of about 1 inHg less for every 1,000 ft of elevation, with about 24 inHg near 5,000 ft. It also warns that the pressure in weather reports is "reduced" to sea level; the true local value is the station pressure. That is the one a gauge reading is relative to.

A worked example shows why this matters. Take a chamber held at 33 mbar absolute, which is what a reading of −98 kPa corresponds to on a standard 1013.25 mbar day. Our arithmetic, using NIST's conversion factors (1 inHg = 3.386389 kPa):

Where and when Local atmosphere Gauge reading for 33 mbar abs
Standard sea-level day 1013 mbar −98.0 kPa
Low-pressure day at sea level 1000 mbar −96.7 kPa
About 5,000 ft (about 24 inHg station pressure) about 813 mbar about −78 kPa

The chamber is the same in every row; only the reference changed. If you compare a reading with a datasheet, or one site with another, convert to absolute with the station pressure, or measure absolute in the first place.

Step 2: know which vacuum range you are in

No single gauge covers the whole vacuum scale. Leybold (read 2026-11-21) notes that the pressures measured in vacuum technology span from 2000 mbar down to 10−12 mbar, over 15 orders of magnitude, so each gauge type covers a band of a few decades. The ranges below are the ones Leybold uses to explain which gauges to use; other sources draw the lines slightly differently (Tawi's guide, for example, ends rough vacuum at 25 Torr, about 33 mbar).

Range Pressure (absolute) Gauges normally used
Rough vacuum atmosphere to 1 mbar Direct (force) gauges: Bourdon, capsule, diaphragm, piezo-resistive, capacitance
Medium vacuum 1 to 10−3 mbar Thermal-conductivity (Pirani) or low-range capacitance gauges
High and ultra-high vacuum below 10−3 mbar Ionization gauges (cold or hot cathode)

Our oil-free piston vacuum pumps work entirely in rough vacuum: our deepest rating, −98 kPa, is about 33 mbar absolute. So for most equipment built around such pumps, the question is not which exotic gauge to buy, but which direct gauge, of what accuracy, fitted where.

The vacuum gauge types, and what each one tells you

Gauges fall into two families. Leybold and Pfeiffer Vacuum (both read 2026-11-21) make the same split. Direct gauges measure pressure as force per area, so their reading does not depend on the type of gas. Indirect gauges measure a property of the gas that changes with pressure, such as how well it carries heat or how easily it ionizes, so their reading does depend on the gas.

Gauge How it works Reads Gas-dependent? Where it is used (per the sources)
U-tube manometer Liquid column rises on the vacuum side Gauge if one leg is open to the air; absolute if that leg is closed and evacuated No Simple direct reading in rough vacuum (Tawi)
Bourdon tube dial A curved tube flexes as the outside air pushes on it; gears move a needle Gauge (relative to atmosphere) No Pfeiffer: 1000 to 1 hPa, limited accuracy, read only at the gauge. Leybold: accurate only to about 10 mbar unless the change in atmospheric pressure is corrected
Capsule (aneroid) gauge A sealed, evacuated capsule expands as pressure falls Absolute No Leybold: linear scale, accurate to about 10 mbar, least accurate at the low-pressure end
Diaphragm gauge with extended low range Corrugated diaphragm against an evacuated reference Absolute No Leybold's example reads 1 to 20 mbar to about 0.3 mbar
Piezo-resistive diaphragm sensor Strain resistors on a diaphragm over a reference vacuum Absolute (electrical output) No Pfeiffer: simple, very robust, insensitive to gas inrush, high accuracy
Capacitance diaphragm gauge Diaphragm deflection changes a capacitance Absolute (electrical output) No Each sensor covers about three to four decades (Leybold says three, Pfeiffer four); the family reaches 10−4 to 10−5 mbar
Pirani (thermal conductivity) A heated wire loses heat to the gas Absolute Yes Pfeiffer: 10 hPa down to 10−4 hPa; little or no pressure signal above roughly 10 to 100 hPa, depending on the gas
Cold and hot cathode ionization Gas is ionized; the ion current is measured Absolute Yes Pfeiffer: below about 10−3 hPa. Typical accuracy ±10 % (hot cathode), ±25 % (cold cathode)

1 hPa = 1 mbar.

Why gas type matters for indirect gauges

Indirect gauges are calibrated on air or nitrogen. Leybold explains that for other gases or vapours a correction factor is needed, and Pfeiffer shows that a Pirani reads close to linear for air and nitrogen but deviates clearly for light gases such as helium and heavy ones such as argon. Pfeiffer adds that the correction factors in the manuals only hold over the linear part of the curve. In practice: if a Pirani gauge on a drying or degassing process shows an odd number, the gas mix may be the reason, not the vacuum. A direct gauge on the same port will tell you.

Why a Pirani gauge is the wrong tool near atmosphere

Pfeiffer notes that from roughly 10 hPa up to atmosphere (the limit depends on the gas) the gas's thermal conductivity no longer changes with pressure, so a Pirani has little to go on. That is also the range where a piston vacuum pump spends all its working life. For rough vacuum use a direct gauge. If one instrument has to cover both rough and lower vacuum, Pfeiffer describes combination sensors that put two sensors in one measuring cell for a broader range.

Step 3: check how accurate the gauge is at the end of its scale

Dial gauges are usually specified as a percentage of span, not of the reading; both WIKA examples below are. That is harmless in the middle of the scale and decisive near full vacuum, where the absolute pressure left in the system is small. Two examples from WIKA's published specifications (read 2026-11-21):

  • A general-purpose Bourdon gauge (model 111.10, vacuum ranges to 30 inHg / −1 bar) is specified at ±3/2/3 % of span, ASME B40.100 Grade B.
  • A 100 mm process gauge (data sheet PM 02.15) is specified at EN 837-1 Class 1.0, which WIKA lists as ±1 % of span, ASME B40.100 Grade 1A. It also lists an extra temperature error of up to ±0.4 % of full scale per 10 °C away from the 20 °C reference.

Now apply that to a −1 to 0 bar gauge (span 100 kPa) and a pump rated at −98 kPa, which is about 33 mbar absolute on a standard day. Our calculation:

Gauge Allowed error (of a 100 kPa span) A reading of −98 kPa means the true absolute pressure is somewhere between
Bourdon dial, ±3 % of span (the end-of-scale figure, which applies near full vacuum) ±3 kPa (±30 mbar) about 3 and 63 mbar abs
Bourdon dial, Class 1.0 (±1 % of span) ±1 kPa (±10 mbar) about 23 and 43 mbar abs
Same Class 1.0 dial used 10 °C away from 20 °C ±1.4 kPa about 19 and 47 mbar abs

The weather comes on top of this for any gauge that reads relative to the atmosphere. Leybold's warning that a Bourdon reading is only good to about 10 mbar uncorrected is the same point. So a dial is fine for "is it pulling" checks and for setting suction-cup or clamp vacuum in the −50 to −90 kPa range. It cannot confirm the last few kPa of a pump's rating. To check a rating near full vacuum, or to compare two pumps, use an absolute diaphragm gauge (capsule, piezo-resistive or capacitance) with a range that ends close to the pressures you expect. Leybold notes that a gauge with a low range, say up to 100 mbar, reads low pressures well but cannot show higher ones; on a linear scale you trade range for resolution.

Leybold also warns that gauges are stretched to cover the widest possible range, and that measurement uncertainty rises very rapidly at the upper and lower limits, in some cases by up to 100 %. Read the range the maker gives for accurate measurement, not only the catalogue range.

Step 4: fit the gauge where the pressure really is

In rough vacuum gas flows, and flowing gas needs a pressure difference. Leybold points out that in the laminar-flow region (roughly from atmosphere down to 0.1 mbar) the pressure just in front of the pump is lower than in the vessel, and that even components with a high conductance create such a gradient. Pfeiffer lists the choice of installation location, because of pressure gradients, first among its application notes. In practice:

  1. Measure at the point that matters. For holding force, measure at the cup or the fixture; for a process, at the chamber. A gauge on the pump inlet shows the best vacuum in the system, not the vacuum at the work.
  2. Measure at the pump inlet to judge the pump. With the inlet closed this is the ultimate-vacuum test, described step by step in how to test a vacuum pump. The gap between the reading at the pump and at the work tells you how much the lines, valves and filters cost you.
  3. Do not put the gauge on a thin, long line. Leybold notes that if the connecting line's conductance is too small, it is evacuated too slowly and the gauge shows too high a pressure. Use a short, full-bore tee, and wait until the reading has stopped moving before you write it down.
  4. Keep connections short and seals few. Pfeiffer recommends this because surfaces and seals give off gas, which makes the gauge read high. The effect matters most at low pressures, but every extra joint is also a possible leak.
  5. Mind vibration. Leybold notes that mechanical gauges are somewhat sensitive to vibration, although small vibrations from a direct connection to a backing pump are generally not harmful. If a needle shakes, mount the gauge off the pump on a short line.

Step 5: record the reading so someone else can use it

A vacuum number without its context cannot be compared. With every reading, write down:

  • the value and its unit, marked gauge or absolute (kPa g, mbar abs, inHg vacuum);
  • the gauge type, range and accuracy class, and its last calibration date;
  • where the gauge was fitted (pump inlet, chamber, cup) and whether the inlet was open or closed;
  • for gauge readings, the local station pressure (not the sea-level value from a weather report) and the altitude;
  • the ambient temperature, and for a pump, the supply voltage and how long it had run.

Gauges drift. Pfeiffer notes that vacuum gauges wear and become contaminated and need replacing from time to time, and that no general interval can be given. For readings that go into acceptance documents, use a gauge with a calibration certificate. Accredited laboratories calibrate vacuum gauges from other makers too; VACUUBRAND, for example, states that its DAkkS-accredited laboratory issues certificates from 1300 to 10−3 mbar absolute. For leak-rate acceptance, see our guide to vacuum system leak rate testing.

Which vacuum gauge should you use? A quick guide

Job Suitable gauge
Is the system pulling? Setting cup, clamp or lifter vacuum Bourdon dial, −1 to 0 bar or 0 to 30 inHg, on a tee near the work
Holding force on site, where the weather and altitude change Gauge (relative) reading, ideally a differential transducer open to atmosphere (Validyne)
Checking a pump against its rating, comparing pumps, a process that depends on absolute pressure Absolute diaphragm gauge: capsule, piezo-resistive or capacitance, ranged for the pressures you expect
Machine control (switch points, interlocks, logging) Electronic diaphragm sensor with an analogue or digital output. Pfeiffer describes "active" gauges, typically 24 V DC with a 0 to 10 V signal
Below about 10 mbar Pirani or low-range capacitance gauge; correct for gas type with a Pirani
Below about 0.001 mbar Ionization gauge (outside the range of piston pumps)

For how a controller uses that signal to hold a setpoint, see how to control vacuum pressure.

Measuring the vacuum of HCEM oil-free piston vacuum pumps

We make oil-free piston vacuum pumps for equipment makers. Our parameter tables give the vacuum as gauge pressure, in kPa with mmHg alongside, in two rows: a maximum open-air flow with its vacuum figure, and a maximum vacuum with its flow figure. On our product pages these are described as two configurations of each size, one built for more flow and one for deeper vacuum; ask us which one a given part number is. The absolute column is our conversion of the maximum-vacuum row at a standard atmosphere of 1013.25 mbar; the last column applies the Class 1.0 example above. The tables do not state the atmospheric pressure at which the figures were measured; ask us for the test conditions and the curve if you need them.

Model Max open air flow row (flow; vacuum) Max vacuum row (flow; vacuum) Max vacuum as absolute (our conversion) What a Class 1.0 dial could confirm at that point
HC80D (24 V DC) 28 L/min; −86 kPa 12 L/min; −95 kPa (713 mmHg) about 63 mbar about 53 to 73 mbar abs
HC100D (24 V DC) 50 L/min; −90 kPa 25 L/min; −98 kPa (735 mmHg) about 33 mbar about 23 to 43 mbar abs
HC280A (220 V/50 Hz) 90 L/min; −92 kPa 45 L/min; −98 kPa (735 mmHg) about 33 mbar about 23 to 43 mbar abs
HC580A (220 V/50 Hz) 120 L/min; −92 kPa 60 L/min; −98 kPa (735 mmHg) about 33 mbar about 23 to 43 mbar abs
HC1500D (24 V DC) 240 L/min; −92 kPa 130 L/min; −98 kPa (735 mmHg) about 33 mbar about 23 to 43 mbar abs

So, to check a pump against its maximum-vacuum figure, an ordinary dial can only tell you whether it is in the right region. Use an absolute diaphragm gauge, fit it straight on the pump inlet, and convert with the station pressure on the day. The AC version HC280B and HC580B (110 V/60 Hz), and 12 V and 48 V versions of the DC models, carry the same vacuum figures in our tables; the full range is on our vacuum pump page. Why the open-flow point matters more than the maximum vacuum for most machines is explained in ultimate pressure vs working vacuum, and how altitude affects the pump itself in vacuum pump altitude and ambient conditions.

Common questions

How do you measure vacuum pressure?

With a vacuum gauge connected to the system where the pressure matters. In rough vacuum use a direct gauge (Bourdon, capsule, piezo-resistive or capacitance diaphragm); at lower pressures a Pirani gauge, and below about 0.001 mbar an ionization gauge. State whether the reading is gauge or absolute.

What instrument is used to measure vacuum?

A vacuum gauge. The common types are the U-tube manometer, the Bourdon tube dial, diaphragm gauges (capsule, piezo-resistive and capacitance), Pirani thermal-conductivity gauges and ionization gauges. Each covers only part of the vacuum range.

What units is vacuum measured in?

Gauge units such as −kPa, −bar or inches of mercury (inHg) of vacuum, and absolute units such as mbar, hPa, Pa or Torr. 1 Torr = 133.3224 Pa and 1 inHg = 3.386389 kPa (NIST). Conversions and the formulas between gauge and absolute are in our gauge pressure guide.

Why does my vacuum gauge read differently from the pump's datasheet?

Usually one of four reasons. The atmosphere differs from the test day or you are at altitude (for a gauge reading). The gauge's error, as a percentage of span, is large near full vacuum. The gauge is not on the pump inlet, so line losses or leaks show up. Or the pump is worn. Rule out the first three before blaming the pump.

Can you measure vacuum without a gauge?

Not as a number. You can see that a system holds vacuum, for example from a cup that holds a part, but you cannot tell the level. A U-tube manometer is the simplest real instrument and needs no power.

Building vacuum measurement into your equipment, or choosing the pump behind it? We supply oil-free piston vacuum pumps to equipment makers and distributors. Custom voltage and logo start at 50 units, and qualified OEM customers and distributors can apply for samples (subject to review), which take 10 to 20 days. See the OEM and ODM page, or send us your target vacuum, where you will measure it, and the flow you need. Our engineers will tell you which pump fits and what data we have at your working pressure.