Which Compressor for an Oxygen Concentrator? Piston vs Scroll vs Diaphragm, and the Specs OEMs Must Set

HCEM Engineering Team

Most stationary oxygen concentrators use an oil-free piston compressor, because it delivers 15 to 150 L/min at the 1.4 to 2.5 bar that the sieve beds need, in a compact and robust package. Scroll compressors run smoother and quieter at the source but cost more and are rarely made this small; diaphragm compressors are the cleanest but move little air for their size, so they suit small portable units. Whichever type you choose, the concentrator only works as well as the compressor's flow at bed pressure, its air quality, its heat and its noise.

This guide explains what the pressure swing adsorption (PSA) beds need from the compressor, compares the three designs on those points, works through the sizing maths and ends with a specification checklist. We make oil-free piston compressors, so read the comparison knowing that. The examples use datasheet values from our own range.

What does the compressor do inside an oxygen concentrator?

A medical oxygen concentrator separates oxygen from room air. The compressor pushes air into a bed of zeolite molecular sieve. Under pressure the zeolite holds back nitrogen, and oxygen-rich gas passes through to a product tank. Before the bed is full, a valve switches the feed to a second bed, and the first is vented so the nitrogen leaves. The cycle repeats every few seconds, day and night.

That process sets six demands on the compressor:

  • Flow at bed pressure. The product flow and purity depend directly on how much air reaches the beds each cycle.
  • Pressure. Typically 1.4 to 2.5 bar gauge (20 to 36 psi) for stationary PSA units, often lower in portables.
  • Oil-free air. Oil damages the sieve, and the product gas goes to a patient.
  • Low noise and vibration. The device runs beside a bed, often all night.
  • Heat. Hot feed air lowers the sieve's capacity, and the cabinet has to get rid of the motor's heat.
  • Continuous duty on the right supply. Mains AC for home and hospital units, battery DC for portables.

How much air does an oxygen concentrator compressor need to deliver?

Start from the oxygen you need and work back to the air. Air is 20.9% oxygen. The sieve beds recover only part of that oxygen into the product; the rest leaves with the nitrogen during venting. That share is the oxygen recovery, and in small medical PSA units it is often around 30 to 40%, depending on the sieve, the bed design and the cycle timing. The feed air the compressor must deliver is:

Feed air (L/min) = product flow × product purity ÷ (0.209 × recovery)

Worked example, 5 L/min unit. Product 5 L/min at 93% oxygen, recovery 35%. Oxygen in the product: 5 × 0.93 = 4.65 L/min. Oxygen the feed must carry: 4.65 ÷ 0.35 = 13.3 L/min. Air needed: 13.3 ÷ 0.209 ≈ 63.6 L/min (2.25 cfm). Add about 10% for ring wear, high room temperature and low mains voltage, and the compressor should deliver about 70 L/min at the bed pressure.

Recovery moves the answer a long way, so measure it on your own beds before you fix the compressor:

Product flow at 93% O₂ Feed air at 30% recovery At 35% recovery At 40% recovery
1 L/min 14.8 L/min 12.7 L/min 11.1 L/min
3 L/min 44.5 L/min 38.1 L/min 33.4 L/min
5 L/min 74.2 L/min 63.6 L/min 55.6 L/min
10 L/min 148.3 L/min 127.1 L/min 111.2 L/min

A useful rule of thumb follows from the table: the compressor moves roughly 11 to 15 litres of air for every litre of oxygen the patient receives. For a 3 L/min unit at 35% recovery, 38.1 L/min plus 10% margin is about 42 L/min (1.48 cfm). For a 10 L/min unit the same assumptions give 127 L/min (4.49 cfm), which is why 10 L concentrators need a much larger compressor (for example our HC580DC, ≥140 L/min at 200 kPa, 12/24/48 V DC), or two in parallel. To switch between L/min and cfm, use our CFM to L/min converter, and see what CFM, SCFM and ACFM mean before comparing figures from different suppliers.

What pressure do the sieve beds need, and why does it vary?

A PSA bed does not sit at one pressure. Each time a bed starts its feed step, its pressure rises from near atmospheric towards the cycle's peak, then drops again when it vents. The compressor therefore works against a back-pressure that swings every few seconds. Higher peak pressure lets the zeolite hold more nitrogen, but costs compressor flow, power and heat.

Two things follow for the specification:

  • Ask for the flow curve, not one point. A single rating such as "50 L/min at 1.4 bar" tells you the flow near the top of the cycle. The flow across 0.5 to 2.5 bar tells you how the compressor behaves over the whole cycle.
  • Match the rating pressure to your cycle. Our oxygenation compressors are rated at either 140 kPa (1.4 bar, 20 psi) or 200 kPa (2 bar, 29 psi). Compare suppliers at the same pressure, or the numbers mean little. Our guide to pulsation and airflow stability covers how to test this on a bench.

Piston, scroll or diaphragm: how do they compare?

All three can deliver oil-free air. They differ in how they seal the compression space, and that decides flow per size, smoothness, noise and wear.

Point Oil-free piston Oil-free scroll Diaphragm
How it compresses Piston with a dry, self-lubricating ring and reed valves Orbiting spiral inside a fixed spiral, tip seals, no valves Flexing membrane with valves, no sliding seal
Flow in a compact size High: about 15 to 150 L/min at 1.4 to 2.5 bar Good, but few models are made as small as concentrator duty Low per head; several heads for more flow
Pressure at concentrator duty Comfortable Comfortable Fine for small flows; flow falls quickly with pressure
Pulsation and vibration Pulsating; needs good mounts and a buffer volume Smooth, low vibration Pulsating, light
Noise at the source Moderate; managed with the cabinet Low Low to moderate
Main wear parts Piston rings, valves, bearings Tip seals, bearings Membrane, valves
Cost and supply Lowest cost per L/min; many suppliers Higher cost; fewer small suppliers Low cost in small sizes
Typical concentrator use Stationary 3 to 10 L/min units and larger portables Some premium stationary units Small portable and pulse-dose units

Piston designs dominate stationary concentrators because they give the most flow at bed pressure for their size and cost, and tolerate the pressure cycling well. The cabinet has to handle their pulsation and noise, and ring wear sets the service interval.

Scroll compressors have no valves and compress continuously, so the flow is smooth and the source noise low. The trade-offs are price, few models at 50 to 100 L/min, and tight clearances that dislike heat and dust.

Diaphragm compressors have nothing sliding in the air path, so there is no wear dust. Their flow drops steeply as pressure rises, so a 5 L/min stationary unit would need a large multi-head pump. Membrane fatigue sets the service interval.

Why must the air be oil-free?

Two reasons. First, the zeolite adsorbs oil vapour and aerosol and does not give them back during venting, so oil slowly blocks the sieve and purity falls. Second, the product gas is breathed by a patient, often for hours a day. Filters after an oil-lubricated compressor must never fail; an oil-free compressor removes the oil at the source. For how air purity is classified, see our guide to ISO 8573-1 air purity classes.

Water is the other enemy: zeolite adsorbs it even more strongly than nitrogen, and moisture slowly lowers the beds' capacity. Cool the compressed air and drain condensate before it reaches the beds.

How quiet does the compressor need to be?

Many home concentrators on the market declare roughly 40 to 50 dB(A) for the finished device. Bare compressors are louder than that: the six models in the table further down are rated from ≤49 dB(A) to ≤62 dB(A), depending on size. The gap is closed by the cabinet: a lined enclosure, soft mounts, an intake silencer and a muffler on the sieve-bed vent, whose exhaust pulse is often louder than the compressor itself.

Three pieces of decibel arithmetic help in the design review:

  • Sources add on a log scale. Two compressors at 52 dB(A) each give 10 × log₁₀(2 × 105.2) ≈ 55 dB(A), not 104.
  • Distance matters. In free field, sound pressure falls about 6 dB per doubling of distance. A level measured at 0.3 m is about 10 dB lower at 1 m (20 × log₁₀(1 ÷ 0.3) ≈ 10.5 dB). Always check the measurement distance and conditions behind a datasheet figure.
  • The gap is the target for the cabinet. A compressor at 52 dB(A) in a device aimed at 40 dB(A) needs the enclosure and silencers to remove about 12 dB at the same distance.

Our guide to acoustic enclosures and vibration isolation covers the cabinet side in more detail.

How do you manage heat in a concentrator cabinet?

Nearly all the electrical power a compressor draws ends up as heat, some in the compressed air and most in the motor and cylinder. That heat matters twice: hot feed air lowers the zeolite's capacity, so purity falls on hot days, and the cabinet must carry the heat away.

Worked example. A 5 L/min unit uses a compressor drawing up to 370 W, such as our HC280. To keep the cabinet air no more than 10 K above the room, the cooling airflow must remove 370 W: mass flow = 370 ÷ (1,005 J/kg·K × 10 K) ≈ 0.037 kg/s. At 1.2 kg/m³ that is about 110 m³/h (roughly 1,840 L/min) of cooling air, about 25 times the compressor's own flow. It has to pass the compressor and leave the cabinet without carrying fan noise with it.

For our AC models in the table below (HC210, HC280 and HC580), the datasheets give an ambient range of −5 to 40 °C and a thermal switch at 145 ± 5 °C that protects the motor. Design the cabinet so the compressor stays well inside that range at your highest room temperature, and check the compressor's duty cycle rating against the 24-hour duty a concentrator demands. Our OEM integration guide covers mounting, thermal management and port sizing.

AC or DC: which supply suits stationary and portable units?

Stationary units run from the mains. The UK and Germany supply 230 V at 50 Hz. A single-phase AC compressor runs at a speed set by the mains frequency, so 50 Hz and 60 Hz versions are different builds: our HC280, for example, comes as the HC280A for 220 V/50 Hz (1400 rpm) and the HC280B for 110 V/60 Hz (1700 rpm). Order the 50 Hz build for UK and EU devices, and ask the supplier to confirm performance at your highest and lowest mains voltage.

Portable units run from a battery, usually through a DC compressor whose speed the controller can vary with the setting. Power draw decides runtime and battery size:

  • Worked example. A pack of 98 Wh (just under the common 100 Wh airline limit for lithium-ion batteries) running a 120 W compressor such as the HC120Z-01 lasts 98 ÷ 120 = 0.82 h, about 49 minutes, for the compressor alone at full power. At 13 V it draws 120 ÷ 13 ≈ 9.2 A. The 52 W HC80Z-11 stretches the same pack to about 1.9 h at 4 A.
  • Current sets the wiring. A 360 W compressor draws 15 A at 24 V but 30 A at 12 V, which decides cable size, connectors and fuse rating.

For a fuller comparison, read AC vs DC oil-free air pumps for OEM equipment.

Which HCEM compressors fit which concentrator size?

Here are six of our oil-free oxygenation compressors, with datasheet values. The last column applies the formula above at 93% purity and 35% recovery, before margin. Note the rating pressure: the 140 kPa models will deliver less at a 200 kPa bed peak. It is a first estimate to narrow the choice; your own bed tests decide.

Model Supply Rated flow Power Noise Weight Rough O₂ output at 35% recovery
HC80Z-11 13 V DC ≥14 L/min at 140 kPa ≤52 W ≤49 dB(A) 360 g about 1.1 L/min
HC120Z-01 13 V DC ≥42 L/min at 140 kPa ≤120 W ≤52 dB(A) 1.1 kg about 3.3 L/min
HC210 220 V/50 Hz or 110 V/60 Hz ≥50 L/min at 140 kPa ≤230 W ≤52 dB(A) 4.3 kg about 3.9 L/min
HC280 220 V/50 Hz or 110 V/60 Hz ≥70 L/min at 200 kPa ≤370 W ≤52 dB(A) 6.0 kg about 5.5 L/min
HC280DC 12, 24 or 48 V DC ≥75 L/min at 200 kPa ≤360 W ≤62 dB(A) 6.0 kg about 5.9 L/min
HC580 220 V/50 Hz or 110 V/60 Hz ≥120 L/min at 200 kPa ≤580 W ≤62 dB(A) 10.8 kg about 9.4 L/min

Read it the way you would any datasheet. The HC80Z-11 (88 × 46 × 73 mm) and HC120Z-01 (129 × 70 × 118 mm) are built for battery-powered designs. The HC280 matches the 5 L/min worked example, with about 10% margin at 35% recovery. For a 10 L/min design the HC580 (≥120 L/min) is just short even at 40% recovery once the 10% margin is added (about 122 L/min); at 35% you need about 140 L/min, which is why the recovery figure from your bed tests has to come before the compressor choice. The full range, including other voltages, is in our oxygen concentrator compressor collection.

What should an OEM put in the compressor specification?

Send every supplier the same list, and ask each to answer it in writing:

  1. Flow: minimum flow at your peak bed pressure, plus the flow curve from 0.5 to 2.5 bar, and the test conditions behind it.
  2. Supply: voltage and frequency (230 V/50 Hz for the UK and EU), or the DC voltage and its range from full to flat battery.
  3. Power and current: at your operating point, plus the start-up current.
  4. Noise: dB(A) with the measurement distance and mounting stated.
  5. Heat: outlet air temperature, allowed ambient range, and the motor's thermal protection.
  6. Duty and life: continuous duty at your pressure, the life-test basis, and which parts wear first.
  7. Air path: confirmation that it is oil-free, the materials that touch the air, and the inlet filter.
  8. Mechanical: dimensions, mounting pattern, port sizes and weight, plus the vibration it passes into the chassis.
  9. Documents: the datasheets, test reports and material declarations your device's technical file needs.

On regulation: in the EU an oxygen concentrator is a medical device under Regulation (EU) 2017/745 (MDR), and ISO 80601-2-69 sets the particular safety and performance requirements for concentrator equipment. The device maker carries the conformity assessment; the compressor is a component inside it. Agree early which supplier documents your notified body will want to see. For more on reading medical compressor datasheets, see our portable medical compressor spec guide. If you need a different flow, voltage, port direction or mounting, our OEM and ODM service covers custom builds, and our customer cases show other projects.

Common questions

What air compressor can I use for an oxygen concentrator?

An oil-free compressor rated for continuous duty, at the flow and pressure your sieve beds need. A workshop air compressor is the wrong tool: most are oil-lubricated or built for 8 bar or more with a tank, they are loud, and they are not designed to run 24 hours a day.

Is a scroll compressor better than a piston compressor for an oxygen concentrator?

It is smoother and quieter at the source, but not better on every point. A piston compressor gives more flow at bed pressure for its size and cost, and a good cabinet closes most of the noise gap. Choose scroll when low vibration is worth the extra cost and a model of the right size exists.

Why does oxygen purity drop when the room is hot?

At higher temperature the zeolite holds less nitrogen per cycle, and a hot compressor delivers less air. Both lower purity. Better cabinet cooling, or a heat exchanger on the compressed air, usually restores it.

How long does an oxygen concentrator compressor last?

It depends on pressure, temperature and hours run, so ask each supplier for the life-test conditions, not just a number of hours. In a piston compressor the rings and valves usually wear first; check that they can be replaced and how often.

Browse our oxygen concentrator compressors, or send us your product flow and purity, bed pressure, supply voltage and noise target, and our engineers will suggest a model, including OEM versions built for your concentrator.